V1.2 30/12/10 Page | 8 1048 Technical Manual
3.3. Measure Current
The unit measures current in three selectable ranges, 0 to 0.22mA (displayed in uA), 0 to
2.2mA and 0 to 22mA (plus over-range to 70mA).
SPECIAL NOTE: Although the top range is nominally 22mA span, it can be safely
used to measure current up to 70mA.
Set Function switch to MEAS
Set OFF/V/mA switch to mA
Set Range switch to the desired range
Set Output switch to NORM
Connect the unit to the signal source observing correct polarity, the current will be
displayed with 3.5 or 4.5 resolution - see Section 3.1.1.
Connect the unit to the sensor or
transmitter observing the correct polarity.
The current drawn is displayed.
3.4. Power and calibrate - transmitters/sensors
If the loop’s power supply is not available it is possible to use the unit’s internal 22V supply
to power the loop. Set up the unit as described below.
Set Function switch to SRC
Set the Multi-Turn Output Control fully clockwise.
Set OFF/V/mA Switch to mA
Set Range switch to the desired range
Set Output switch to NORM
Connect the unit to the sensor or transmitter
observing the correct polarity.
The current drawn is displayed.
-
+
-
+
Sensor
Measure & Power Mode
2 w ire transducer/transmitter